Optimizing image quality and minimizing radiation dose in pediatric abdominal multiphase contrast-enhanced computed tomography: a study on CARE kV and CARE Dose 4D

优化儿童腹部多期增强CT图像质量并最大限度降低辐射剂量:CARE kV和CARE Dose 4D研究

阅读:1

Abstract

BACKGROUND: Multiphase contrast-enhanced computed tomography (CECT) is a commonly used modality in pediatric computed tomography (CT) scans. However, the purposes and focus of each phase, such as CT angiography (CTA), and parenchymal phase, are different. In routine practice, the same scanning parameters are used for all phases, resulting in unnecessary radiation exposure for children. Accurately and rapidly adjusting the scanning parameters for each phase of CECT is challenging in clinical settings. This retrospective cross-sectional study was designed to investigate the feasibility of using both CARE kV and CARE Dose 4D to reduce the radiation dose while maintaining diagnostic quality in multiphase CECT scans of children. METHODS: Overall, 57 children (33 males and 24 females) who underwent multiphase abdominal CECT in Xinjiang Hospital of Beijing Children's Hospital with an average age of 6.52±4.30 years (range, 0.1-15 years), were enrolled. The tube voltage was automatically modulated using CARE kV. The tube current was automatically modulated using CARE Dose 4D. Different dose saving optimization indices (DI) were used for the three phases: a DI value of 3 was used for the unenhanced CT phase, a DI value of 12 was used for the CTA phase, and a DI value of 7 was used for the parenchymal phase. The tube voltage and volume CT dose index (CTDI(vol)) were recorded for each phase. Two reviewers subjectively evaluated the overall image quality and noise level of the three phases using a 5-point Likert scale (1-2 points: unqualified, 3 points: qualified, 4 points: better, 5 points: best). The CT and noise values of the descending aorta, liver, and back muscle were measured objectively. The voltage distribution and the image quality and CTDI(vol) in each phase were compared. RESULTS: The most selected tube voltage in the unenhanced CT, CTA, and parenchymal phases was 100 kV (49/57, 85.96%), 70 kV (36/57, 63.16%), and 80 kV (32/57, 56.14%), respectively. The differences between the three phases were statistically significant (P<0.001). The CTDI(vol) values of the three phases were 3.99±1.99, 2.02±1.71, and 3.18±2.10 mGy, respectively, with a significant difference between the three phases (P<0.001). The CTDI(vol) decreased linearly as the DI value increased. All images met the diagnostic requirements. The overall quality scores for the three phases were 4.24±0.42, 4.41±0.49, and 4.50±0.45, respectively, with no significant linear relationship with the change in the DI. CONCLUSIONS: The combined use of CARE Dose 4D and CARE kV could effectively reduce the radiation dose in children during multiphase abdominal CECT without compromising the diagnostic image quality.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。