Evaluation of Reliability of Dynamic Scapholunate Distance Measured on 4D CT-Acquired Images

评估基于4D CT图像测量的动态舟月骨间距的可靠性

阅读:1

Abstract

Purpose  A technique to measure scapholunate distance based on four-dimensional computed tomography (4D CT)-acquired images is presented. Materials and Methods Intra-observer variability was evaluated through a repeated-measures study. A 4D CT of seven patients suspected of scapholunate lesion was performed. Anatomical landmarks were identified on a three-dimensional reconstructed model of the wrist. All 4D CT datasets were evaluated thrice by two observers. Standard deviation of the differences between two measurements, interclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were calculated. Results  Intra-observer variability for the expert observer (ICC > 0.95) was lower than that of the novice observer (ICC > 0.77) and interobserver variability was low (ICC > 0.85). For the expert observer, measurement error (SEM < 0.13 mm and MDC < 0.36 mm) was smaller than that of the novice observer (SEM < 0.45 mm and MDC < 1.24 mm). Both SEM and MDC values were low, compared to the scan resolution and the absolute value of intervals. Conclusion  The proposed assessment results in a reproducible and reliable measurement of scapholunate distance.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。