Clinical implementation of a commercial synthetic computed tomography solution for radiotherapy treatment of glioblastoma

商业合成计算机断层扫描解决方案在胶质母细胞瘤放射治疗中的临床应用

阅读:1

Abstract

BACKGROUND AND PURPOSE: Magnetic resonance (MR)-only radiotherapy (RT) workflow eliminates uncertainties due to computed tomography (CT)-MR image registration, by using synthetic CT (sCT) images generated from MR. This study describes the clinical implementation process, from retrospective commissioning to prospective validation stage of a commercial artificial intelligence (AI)-based sCT product. Evaluation of the dosimetric performance of the sCT is presented, with emphasis on the impact of voxel size differences between image modalities. MATERIALS AND METHODS: sCT performance was assessed in glioblastoma RT planning. Dose differences for 30 patients in both commissioning and validation cohorts were calculated at various dose-volume-histogram (DVH) points for target and organs-at-risk (OAR). A gamma analysis was conducted on regridded image plans. Quality assurance (QA) guidelines were established based on commissioning phase results. RESULTS: Mean dose difference to target structures was found to be within ± 0.7 % regardless of image resolution and cohort. OARs' mean dose differences were within ± 1.3 % for plans calculated on regridded images for both cohorts, while differences were higher for plans with original voxel size, reaching up to -4.2 % for chiasma D2% in the commissioning cohort. Gamma passing rates for the brain structure using the criteria 1 %/1mm, 2 %/2mm and 3 %/3mm were 93.6 %/99.8 %/100 % and 96.6 %/99.9 %/100 % for commissioning and validation cohorts, respectively. CONCLUSIONS: Dosimetric outcomes in both commissioning and validation stages confirmed sCT's equivalence to CT. The large patient cohort in this study aided in establishing a robust QA program for the MR-only workflow, now applied in glioblastoma RT at our center.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。