A new seven-axis robotic-assisted total hip arthroplasty system improves component positioning: a prospective, randomized, multicenter study

一种新型七轴机器人辅助全髋关节置换系统可改善假体定位:一项前瞻性、随机、多中心研究

阅读:1

Abstract

This study compared the radiologic and clinical outcomes of a new seven-axis robotic-assisted total hip arthroplasty (THA) and conventional THA. Hundred and four patients were randomly assigned to two groups-the robotic-assisted THA group (RAS group) and the conventional THA group (CON group). The preoperative and postoperative Harris Hip score (HHS), acetabular inclination, anteversion, femoral offset, and leg length discrepancy (LLD) were compared. During the follow-up, no patients had any complications that could be associated with the use of the robot. The proportion of acetabular cups in the safety zone was significantly higher in the RAS group than that in the CON group. The two groups had significantly different mean absolute difference of inclination and anteversion. There was no significant difference in the postoperative HHSs, changes in HHSs, femoral offset, and lower limb length between the two groups. The seven-axis robotic-assisted THA system is safe and effective, and leads to better acetabulum cup positioning compared to conventional THA. The improvements observed in the HHS, LLD, and femoral offset in the RAS group were similar to those in the CON group.Clinical trial registration time: 19/05/2022.Clinical trial registration number: ChiCTR2200060115.