Analytical performance evaluation of the Mindray enzymatic assay for hemoglobin A(1c) measurement

对迈瑞酶法测定糖化血红蛋白A1c的分析性能进行评价

阅读:2

Abstract

Hemoglobin A(1c) (HbA(1c)) plays a crucial role in diabetes management. We aimed to evaluate the analytical performance of a new enzymatic method kit for HbA(1c) measurement. The performance of the enzymatic method, including precision, accuracy, and linearity, was evaluated. Moreover, the interference effect from conventional interferents, Hb derivatives, Hb variants, and common drugs were assessed. In addition, the agreement of HbA(1c) results was compared between enzymatic methods, cation-exchange high-performance liquid chromatography (HPLC), and immunoassays. The intra-assay, between-assay, and total precision of HbA(1c) were all lower than 2%. HbA(1c) showed good linearity within the range of 3.96-20.23%. The enzymatic assay yielded results consistent with the external quality control samples, with a bias of less than ± 6% from the target values. The enzymatic method showed no interference from bilirubin, intralipid, vitamin C, Hb derivatives, common Hb variants, as well as antipyretic analgesics and hypoglycemic drugs. The HbA(1c) results of the enzymatic assay showed good agreement and accuracy compared to those obtained from the HPLC method and the immunoassay. The enzymatic method kit performed on the BS-600M chemistry analyzer is a reliable and robust method for measuring HbA(1c). It is suitable for routine practice in clinical chemistry laboratories.