Assessment of Vecuronium Quality Using Near-Infrared Spectrometry

使用近红外光谱法评估维库溴铵的质量

阅读:6
作者:James T Isaacs, Philip J Almeter, Aaron N Hunter, Thomas A Lyman, Stephanie P Zapata, Bradley S Henderson, Seth A Larkin, Lindsey M Long, Megan N Bossle, Smaran A Bhaktawara, Matthew F Warren, Austin M Lozier, Joshua D Melson, Savannah R Fraley, Eunice Hazzel L Relucio, Margaret A Felix, Jeffrey W R

Abstract

This study employed Fourier Transform near-infrared spectrometry to assess the quality of vecuronium bromide, a neuromuscular blocking agent. Spectral data from two lots of vecuronium were collected and analyzed using the BEST metric, principal component analysis (PCA) and other statistical techniques. The results showed that there was variability between the two lots and within each lot. Several outliers in the spectral data suggested potential differences in the chemical composition or sample condition of the vials. The outliers were identified and their spectral features were examined. A total of eight unique outliers were found in the PC space from PCs 1 to 9, so 22% of the total vials were outliers. The study findings suggest that the manufacturing process of vecuronium bromide may have been operating outside of a state of process control. Further investigation is needed to determine the source of these variations and their impact on the safety and efficacy of the drug product.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。