Laboratory and field testing of an automated atmospheric particle-bound reactive oxygen species sampling-analysis system

自动化大气颗粒结合活性氧采样分析系统的实验室和现场测试

阅读:5
作者:Yungang Wang, Philip K Hopke, Liping Sun, David C Chalupa, Mark J Utell

Abstract

In this study, various laboratory and field tests were performed to develop an effective automated particle-bound ROS sampling-analysis system. The system uses 2' 7'-dichlorofluorescin (DCFH) fluorescence method as a nonspecific, general indicator of the particle-bound ROS. A sharp-cut cyclone and a particle-into-liquid sampler (PILS) were used to collect PM(2.5) atmospheric particles into slurry produced by a DCFH-HRP solution. The laboratory results show that the DCFH and H(2)O(2) standard solutions could be kept at room temperature for at least three and eight days, respectively. The field test in Rochester, NY, shows that the average ROS concentration was 8.3 ± 2.2 nmol of equivalent H(2)O(2) m(-3) of air. The ROS concentrations were observed to be greater after foggy conditions. This study demonstrates the first practical automated sampling-analysis system to measure this ambient particle component.