Development of a dual-component infection-resistant arterial replacement for small-caliber reconstructions: A proof-of-concept study

开发用于小口径重建的双组分抗感染动脉替代品:一项概念验证研究

阅读:15
作者:Sonia Zia, Adrian Djalali-Cuevas, Michael Pflaum, Jan Hegermann, Daniele Dipresa, Panagiotis Kalozoumis, Artemis Kouvaka, Karin Burgwitz, Sofia Andriopoulou, Alexandros Repanas, Fabian Will, Karsten Grote, Claudia Schrimpf, Sotiria Toumpaniari, Marc Mueller, Birgit Glasmacher, Axel Haverich, Lucrezi

Discussion

The study demonstrated the feasibility of fabricating a dual-component small-caliber graft, comprising a scaffold with sufficient biological and biomechanical functionality, and an electrospun PCL/PEG sleeve with tailored biomechanics and antibiotic release.

Methods

The study investigated the effect of nucleases, as part of the decellularization technique, and two sterilization methods (peracetic acid and γ-irradiation), on the scaffold's biological and biomechanical integrity. It also investigated the effect of different PCL/PEG ratios on the antimicrobial, biological and biomechanical properties of the sleeves. Tibial arteries were decellularized using Triton X-100 and sodium-dodecyl-sulfate.

Results

The scaffolds retained the general native histoarchitecture and biomechanics but were depleted of glycosaminoglycans. Sterilization with peracetic acid depleted collagen IV and produced ultrastructural changes in the collagen and elastic fibers. The two PCL/PEG ratios used (150:50 and 100:50) demonstrated differences in the structural, biomechanical and antimicrobial properties of the sleeves. Differences in the antimicrobial activity were also found between sleeves fabricated with antibiotics supplemented in the electrospinning solution, and sleeves soaked in antibiotics.