Analytical and clinical comparison of two fully automated immunoassay systems for the diagnosis of celiac disease

两种全自动免疫分析系统用于乳糜泻诊断的分析和临床比较

阅读:13
作者:Gabriella Lakos, Gary L Norman, Michael Mahler, Peter Martis, Chelsea Bentow, Debby Santora, Alessio Fasano

Conclusion

The QUANTA Flash CD assays have outstanding clinical performance. Of particular clinical significance, in light of proposals to decrease the absolute necessity of biopsy, was the demonstration that 50% of the QUANTA Flash h-tTG IgA results were >10x ULN.

Methods

A total of 74 biopsy-proven CD patients (2 with IgA deficiency) and 138 controls were tested by both methods.

Objective

Here we compared analytical and clinical performance characteristics of two novel automated assay systems for the detection of celiac disease (CD) specific antibodies: QUANTA Flash (INOVA Diagnostics, Inc.) and EliA (Thermo Scientific).

Results

Sensitivities of QUANTA Flash assays ranged from 35.1% to 90.5% and specificities from 96.4% to 99.3%, while sensitivities for EliA assays ranged from 37.8% to 90.5% (equivocal considered positive) and specificities from 97.1% to 100.0%. Good qualitative agreement was found between all assays. Thirty-four (50.0%) of the 68 QUANTA Flash h-tTG IgA positive results were higher than 10 times the upper limit of normal (ULN). In contrast, only 22.8% of the EliA tTG IgA positive samples were >10x ULN. Seventy-three (98.6%) biopsy-proven CD patients were correctly identified with the QUANTA Flash h-tTG IgA+DGP IgG combination, while 64 (86.5%) and 72 (97.3%) (depending on equivocal range) were identified with the same combination of EliA assays.