Electrochemical precipitation of neptunium with a micro electrochemical quartz crystal microbalance

使用微型电化学石英晶体微天平进行镎的电化学沉淀

阅读:6
作者:Adan Schafer Medina, Gretchen Tibbits, Nathalie A Wall, Cornelius F Ivory, Sue B Clark, Haluk Beyenal

Abstract

Microliter volumes are used in electrochemical detection and preconcentration of radionuclides to reduce the dose received by researchers and equipment. Unfortunately, there is a lack of analysis of radionuclides with coupled electrochemical techniques and microliter volume reactors. The goals of this work are 1) to develop a miniaturized micro-electrochemical quartz crystal microbalance (µeQCM) reactor for use in small volume (50-200 µL) electrogravimetric experiments and 2) to use this reactor to characterize the preconcentration of neptunium on carbon electrodes via electroprecipitation. We successfully deposited neptunium in the new µeQCM reactor and verified its operation. We found that preconcentration of neptunium on carbon coated electrodes was possible by chronoamperometry at -1.6 VAg/AgCl. The mass shift of the resulting precipitate was indicative of the amount of neptunium on the electrode, although the correlation between the mass increase and activity of the preconcentrated material was not linear. Neptunium precipitate reduced electron transfer to the solution as evidenced by the increase in charge transfer resistance compared to bare electrodes.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。