Connecting Across Competencies: Leveraging Best Practices for Processing

跨能力连接:利用最佳实践进行处理

阅读:6
作者:Terra A Kremer, Kaumudi Kulkarni, Christopher Ratanski, Lorraine Floyd, Christopher Anderson

Abstract

The AAMI working group ST/WG 93 is finalizing a standard (AAMI ST98) for the cleaning validation of reusable medical devices based on guidance from the technical information report AAMI TIR30:2011/(R)2016. A number of analytical best practices are being considered for this new standard. Test method suitability for processing cleaning validations historically has been established using one positive control and performing an extraction efficiency. The new cleaning validation standard is proposed to require a change from only one replicate test sample to three when performing method suitability. This change will affect manufacturers; therefore, the value of and consideration for performing these additional replicates requires explanation. This article discusses how variation of validation parameters can affect the accuracy and precision during method suitability testing. Multiple replicates are needed to understand the variability of method extraction and impact on cleaning validations of reusable medical devices.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。