Vortex counter-current chromatography

涡旋逆流色谱

阅读:5
作者:Yoichiro Ito, Zhiyong Ma, Robert Clary, Jimmie Powell, Martha Knight, Thomas M Finn

Abstract

A novel counter-current chromatographic system is developed by mounting a vortex column on a type-I coil planet centrifuge. The column is fabricated from a high-density polyethylene disk (16 cm diameter and 5 cm thick) by making multiple holes of various diameters (3-12.5 mm) each arranged in a circle and connected with narrow transfer ducts. The performance of this vortex column is tested with three different two-phase solvent systems with a broad range in hydrophobicity. The results indicated that the smallest diameter column (3mm diameter, 120 units with 42.8 ml capacity) yielded the best separation with the height equivalent to a theoretical plate of 2 cm compared with 20 cm required by the conventional multilayer coil column of high-speed CCC. By avoiding the use of an Archimedean Screw Force, the system shows a low column pressure which would permit safe operation of a large preparative column without a risk of leakage of solvent and column damage.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。