Improved Dose Response of N-(hydroxymethyl)acrylamide Gel Dosimeter with Calcium Chloride for Radiotherapy

氯化钙 N-(羟甲基)丙烯酰胺凝胶剂量计在放射治疗中的剂量响应改进

阅读:9
作者:Khalid A Rabaeh, Rawan E Al-Tarawneh, Molham M Eyadeh, Issra' M E Hammoudeh, Moneeb T M Shatnawi

Abstract

The impact of calcium chloride (CaCl2) on the performance of N-(hydroxymethyl)acrylamide (NHMA) polymer gel dosimeter is studied in this article. The dosimeter was exposed to doses of up to 10 Gy with radiation beam-energy of 10 MV and dose-rates of 300 cGy/min. The relaxation rate (R2) parameter was utilized to explore the performance of irradiated NHMAGAT gels. The dose response in terms of R2 increased from 0.29 to 0.63 Gy-1·s-1 with increasing calcium chloride concentration from 0 to 1000 mM. The results show no substantial impact of dose-rates as well as radiation energies on NHMAGAT samples. For the steadiness of irradiated NHMAGAT dosimeters, it was found that there is no apparent variation in R2 (less than ±3%; standard deviation) up to 3 days. The overall uncertainty of the gel dosimeter with calcium chloride is 4.96% (double standard deviation, 95% confidence level).