Investigation of surgical adhesives for vocal fold wound closure

声带伤口缝合手术粘合剂的研究

阅读:6
作者:Lei Xi Chen, Maxence Coulombe, Francois Barthelat, Almoaidbellah Rammal, Luc Mongeau, Karen Kost

Conclusion

There was a clear tradeoff between adhesive strength and toxicity. The maximum failure strength in shear or tension of the three surgical adhesives ranked from strongest to the weakest was: 1) Glubran 2, 2) BioGlue, and 3) Tisseel. Tisseel was found to be the least toxic of the three adhesives, whereas Glubran 2 was the most toxic. Level of evidence: NA Laryngoscope, 129:2139-2146, 2019.

Methods

Shear and tension tests were performed on 150 porcine larynges. The cytotoxicity of the adhesives to immortalized human vocal fold fibroblasts was investigated using neutral red uptake assays.

Results

The average shear adhesive strength for Tisseel, BioGlue, and Glubran 2 was 13.86 ± 5.03 kilopascal (kPa), 40.92 ± 17.94 kPa, and 68.79 ± 13.29 kPa, respectively. The tensile adhesive strength for Tisseel, BioGlue, and Glubran 2 was 10.70 ± 6.42 kPa, 34.27 ± 12.59 kPa, and 46.67 ± 12.13 kPa, respectively. The vocal fold cell viabilities in extracts of Tisseel, BioGlue, and Glubran 2 were 99.27%, 43.05%, and 1.79%, respectively.