Evaluation of the increase in permeability of the blood-brain barrier during tumor progression after pulsed focused ultrasound

评估脉冲聚焦超声治疗后肿瘤进展过程中血脑屏障通透性的增加

阅读:7
作者:Feng-Yi Yang, Hsin-Ell Wang, Guan-Liang Lin, Hui-Hsien Lin, Tai-Tong Wong

Conclusion

This method could allow enhanced synergistic effects with respect to other brain tumor treatment regimens.

Methods

F98 glioma-bearing Fischer 344 rats were injected intravenously with Evans blue with or without blood-tumor barrier disruption induced by transcranial pulsed high-intensity focused ultrasound. Sonication was applied at a frequency of 1 MHz with a 5% duty cycle and a repetition frequency of 1 Hz. The permeability of the blood-brain barrier was assessed by the extravasation of Evans blue. Contrast-enhanced magnetic resonance images were used to monitor the gadolinium deposition path associated with transcranial pulsed high-intensity focused ultrasound, and the influencing size and location was also investigated. In addition, whole brain histological analysis was performed. The

Purpose

The purpose of this study was to evaluate the permeability of the blood-brain barrier after sonication by pulsed high-intensity focused ultrasound and to determine if such an approach increases the tumor:ipsilateral brain permeability ratio. Materials and

Results

The accumulation of Evans blue in brains and the tumor:ipsilateral brain permeability ratio of Evans blue were significantly increased after pulsed high-intensity focused ultrasound exposure. Evans blue injection followed by sonication showed an increase in the tumor:ipsilateral brain ratio of the target tumors (9.14:1) of about 2.23-fold compared with the control tumors (x4.09) on day 6 after tumor implantation. Magnetic resonance images showed that pulsed high-intensity focused ultrasound locally enhances the permeability of the blood-tumor barrier in the glioma-bearing rats.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。