Conclusions

Large exposure ranges make CBCT doses difficult to generalize. Use of DHP as a metric for estimating effective dose warrants further investigation.

Methods

The English literature on effective dose was reviewed. Data from these studies together with additional data for nine CBCT units were analysed. Descriptive statistics, ANOVA and paired analysis are used to characterize the data.

Results

PubMed and EMBASE searches yielded 519 and 743 publications, respectively, which were reduced to 20 following review. Reported adult effective doses for any protocol ranged from 46 to 1073 µSv for large fields of view (FOVs), 9-560 µSv for medium FOVs and 5-652 µSv for small FOVs. Child effective doses from any protocol ranged from 13 to 769 µSv for large or medium FOVs and 7-521 µSv for small FOVs. Effective doses from standard or default exposure protocols were available for 167 adult and 52 child exposures. Mean adult effective doses grouped by FOV size were 212 µSv (large), 177 µSv (medium) and 84 µSv (small). Mean child doses were 175 µSv (combined large and medium) and 103 µSv (small). Large differences were seen between different CBCT units. Additional low-dose and high-definition protocols available for many units extend the range of doses. DHP was found to reduce average absolute error for calculation of dose by 45% in comparison with DAP. Conclusions: Large exposure ranges make CBCT doses difficult to generalize. Use of DHP as a metric for estimating effective dose warrants further investigation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。