Conclusion
The Immunalysis HEIA K2 Spice kit required no sample preparation, had a high-throughput, and acceptable sensitivity, specificity and efficiency, offering a viable method for screening synthetic cannabinoids in urine that cross-react with JWH-018 N-pentanoic acid antibodies.
Methods
2443 authentic urine samples were analyzed with the HEIA that targets JWH-018 N-pentanoic acid, and a validated LC-MS/MS method for 29 synthetic cannabinoids and metabolites. Semi-quantitative HEIA
Results
Sensitivity, specificity, and efficiency of the immunalysis HEIA K2 Spice kit with the manufacturer's recommended 10 μg/L cutoff were 75.6%, 99.6% and 96.8%, respectively, as compared to the reference LC-MS/MS method with limits of detection of 0.1-10 μg/L. Performance at 5 μg/L was 92.2%, 98.1% and 97.4%, and for the 20 μg/L cutoff were 62.9%, 99.7% and 95.4%. Semi-quantitative results for in-house prepared standards were obtained from 2.5-30 μg/L, and documented acceptable linearity from 5-25 μg/L, with inter-day imprecision <30% (n = 17). Thirteen of 74 synthetic cannabinoids evaluated were classified as highly cross-reactive (≥50% at 10 μg/L); 4 showed moderate cross-reactivity (10-50% at 10 μg/L), 30 low cross-reactivity (<10% at 500 μg/L), and 27 <1% cross-reactivity at 500 μg/L. There was no interference from 102 investigated compounds. Only a mixture containing 1000 μg/L each of buprenorphine/norbuprenorphine produced a positive result above our proposed cutoff (5 μg/L) but below the manufacturer's recommended cutoff concentration (10 μg/L).