A High-Sensitivity Method Based on Advanced Optical Waveguide Technology to Detect HBsAg

基于先进光波导技术的高灵敏度 HBsAg 检测方法

阅读:4
作者:Pingping Xiao, Xiaoxiong Hu

Abstract

A novel method for the detection of the hepatitis B surface antigen (HBsAg) at low concentrations, using the ultrahigh-order guided mode acting as the probe excited by a symmetrical metal-cladding waveguide, is proposed. The method using the fact of the minimum value of the absorption peaks is proportional to the concentration of the sample to be detected to realize the detection of the hepatitis B virus at extremely low concentrations. It is realized that the low concentration of the HBsAg measurement relied on the principle of the minimum value of the absorption peak and the concentration having a good linear relationship. The measurement results indicate that this new method can precisely detect HBsAg at the concentrations in the lower region of the clinical gray area (i.e., below 20 ng/mL), the lower region of the current clinical gray area of HBsAg (below 20 ng/ml) can be measured, and the resolution can be reached (2 ng/mL).