Evaluation of in vitro hemolysis and platelet activation of a newly developed maglev LVAD and two clinically used LVADs with human blood

评估新型磁悬浮 LVAD 和两种临床使用的 LVAD 的体外溶血和血小板活化作用

阅读:7
作者:Zachary B K Berk, Jiafeng Zhang, Zengsheng Chen, Douglas Tran, Bartley P Griffith, Zhongjun J Wu

Abstract

In vitro hemolysis testing remains one of the most important performance measures to judge the hemocompatibility of a left ventricular assist device (LVAD). Clinically relevant operating conditions and appropriate testing blood are essential to infer in vitro data for potential clinical use. This in vitro study was carried out to evaluate and compare the hemolytic performance of a newly developed magnetically levitated (maglev) LVAD (CH-VAD) with two clinically used LVADs (HVAD and HeartMate II (HMII)) using fresh human blood. A small volume (~300 mL) in vitro circulating flow loop was constructed with a LVAD generated flow of 4.5 L/min at the nominal or reported clinical operating speed for each LVAD. The blood was circulated in the loop for 4 hours with samples drawn at baseline and hourly. Plasma-free hemoglobin (PFH) concentrations in the hourly blood samples were determined with spectrophotometry. Normalized index of hemolysis (NIH) was calculated to compare the hemolytic performance of the CH-VAD and the two reference LVADs. Platelet activation was measured with flow cytometry. The experimental test for each device was repeated at least 7 times. The data from this study showed that all the three LVADs generated very low hemolysis (NIH <0.01 g/100 L). The CH-VAD was found to have a significantly lower NIH value (0.00135 ± 0.00032 g/100 L) compared to the HVAD (0.00525 ± 0.00183 g/100 L) and the HMII (0.00583 ± 0.00182 g/100 L). No statistically significant difference in device-generated hemolysis was found between the HVAD and the HMII. The level of platelet activation induced by the CH-VAD is significantly lower than those by the HVAD and the HMII. The data suggest that the shear-induced hemolysis and platelet activation of the CH-VAD are acceptable relative to the two LVADs currently in clinical use.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。