Performance of MassARRAY system for the detection of SARS-CoV-2 compared to real-time PCR

MassARRAY 系统检测 SARS-CoV-2 的性能与实时 PCR 的比较

阅读:6
作者:Fatimah AlMutawa, Ana Cabrera, Feifei Chen, Johan Delport

Background

Early identification of COVID-19 (coronavirus disease of 2019) by diagnostic tests played an important role in the isolation of infectious patients and management of this pandemic. Various methodologies and diagnostic platforms are available. The current "gold standard" for SARS-CoV-2 (severe acute respiratory syndrome coronavirus 2) diagnosis is real-time reverse transcriptase-polymerase chain reaction (RT-PCR). To overcome the limitations posed by the short supply experienced early during the pandemic and to increase our capacity, we assessed the performance of the MassARRAY System (Agena Bioscience).

Conclusion

Our study demonstrates that the MassARRAY System is an accurate and sensitive method for SARS-CoV-2 detection. Following the discordant agreement with an alternate RT-PCR test, the performance was found to have sensitivity, specificity, and accuracy exceeding 97%, making it a viable diagnostic tool. It can be used as an alternative method during periods when real-time RT-PCR reagent supply chains are disrupted.

Methods

MassARRAY System (Agena Bioscience) combines RT-PCR (reverse transcription-polymerase chain reaction) with high-throughput mass spectrometry processing. We compared the MassARRAY performance to a research-use-only E-gene/EAV (Equine Arteritis Virus) assay and RNA Virus Master PCR. Discordant

Results

186 patient specimens were analyzed using the MassARRAY SARS-CoV-2 Panel. The performance characteristics were as follows: the positive agreement was 85.71%, 95% CI (78.12 - 91.45), and the negative agreement was 96.67%, 95% CI (88.47 - 99.59). 19/186 (10.2%) results were found to be discordant and assessed by a different assay with the exception of 1, where the sample was not available for repeat testing. 14 out of 18 agreed with the MassARRAY after testing with the secondary assay. The overall performance after discordance testing was as follows: the positive agreement was 97.3%, 95% CI (90.58 - 99.67), and the negative agreement was 97.14%, 95% CI (91.88 - 99.41).