Conclusion
Calculating the ED50 and EDref provides an effective way to compare values from different platelet function assays. Flow cytometry enables correct platelet function testing as long as platelet count is > 30 x109/L whereas impedance aggregometry and in-vitro bleeding time are inconsistent unless platelet count is > 100 x109/L.
Methods
We assessed platelet function in whole blood samples of healthy volunteers at decreasing platelet counts (> 100, 80-100, 50-80, 30-50 and < 30 x109/L) using two POC tests: impedance aggregometry and in-vitro bleeding time. Flow cytometry served as the gold standard. The number of platelets needed to reach 50% of the maximum function (ED50) and the lower reference limit (EDref) were calculated to define limits of test validity.
Results
The minimal platelet count required for reliable test results was 100 x109/L for impedance aggregometry and in-vitro bleeding time but only 30 x109/L for flow cytometry. Comparison of ED50 and EDref showed significantly lower values for flow cytometry than either POC test (P value < 0.05) but no difference between POC tests nor between the used platelet agonists within a test method.