Highly Sensitive Uric Acid Detection Based on a Graphene Chemoresistor and Magnetic Beads

基于石墨烯化学电阻和磁珠的高灵敏度尿酸检测

阅读:9
作者:Wangyang Zhang, Xiaoqiang Zhao, Lina Diao, Hao Li, Zhonghao Tong, Zhiqi Gu, Bin Miao, Zhan Xu, Han Zhang, Yue Wu, Jiadong Li

Abstract

In this study, we developed a low-cost, reusable, and highly sensitive analytical platform for the detection of the human metabolite uric acid (UA). This novel analysis platform combines the graphene chemoresistor detection technique with a magnetic bead (MB) system. The heterojunction (single-layer graphene and HfO2 thin-film material) of our graphene-based biosensor worked as a transducer to detect the pH change caused by the specific catalytic reaction between UA and uricase, and hence acquires a UA concentration. Immobilization of uricase on MBs can decouple the functionalization steps from the sensor surface, which allows the sensor to be reusable. Our microsensor platform exhibits a relatively lower detection limit (1 μM), high sensitivity (5.6 mV/decade), a linear range (from 1 μM to 1000 μM), and excellent linearity (R2 = 0.9945). In addition, interference assay and repeatability tests were conducted, and the result suggests that our method is highly stable and not affected by common interfering substances (glucose and urea). The integration of this high-performance and compact biosensor device can create a point-of-care diagnosis system with reduced cost, test time, and reagent volume.