Assessment of ELISA-based method for the routine examination of serum indoxyl sulfate in patients with chronic kidney disease

慢性肾脏病患者血清硫酸吲哚美辛常规检测ELISA方法的评价

阅读:11
作者:Shuangdi Duan, Jiayi Pi, Chun-Hsiang Wang, Yi-Chou Hou, Chung-Ying Andy Lee, Cheng-Jui Lin, Liyang Shi, Kung-Chia Young, Hung-Yu Sun3

Conclusions

The Leadgene ELISA kit exhibits comparable efficacy to UPLC-MS/MS in quantifying serum IS levels, supporting that ELISA would be a personalized method for monitoring the dynamic changes in serum IS levels in dialysis patients to prevent the progression of CKD.

Methods

The efficiencies of three commercial ELISA kits in determination of serum IS were validated by comparing with ultra-performance liquid chromatography (UPLC)-MS/MS-based method using Bland-Altman analysis. The associations between kidney parameters and serum IS levels determined by ELISA kit from Leadgene and UPLC-MS/MS were evaluated by Spearman correlation coefficient in a CKD validation cohort.

Results

ELISA kit from Leadgene showed clinical agreement with UPLC-MS/MS in the determination of serum IS levels (p = 0.084). In patients with CKD, Spearman's correlation analysis revealed a perfect correlation between the IS levels determined using the Leadgene ELISA kit and UPLC-MS/MS (r = 0.964, p < 0.0001). IS levels determined using the Leadgene ELISA kit were associated with the estimated glomerular filtration rate (r = -0.772, p < 0.0001) and serum creatinine concentration (r = 0.824, p < 0.0001) in patients with CKD, and on dialysis (r = 0.557, p = 0.006). Conclusions: The Leadgene ELISA kit exhibits comparable efficacy to UPLC-MS/MS in quantifying serum IS levels, supporting that ELISA would be a personalized method for monitoring the dynamic changes in serum IS levels in dialysis patients to prevent the progression of CKD.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。