Influence of High-Intensity Focused Ultrasound (HIFU) Ablation on Arteries: Ex Vivo Studies

高强度聚焦超声 (HIFU) 消融对动脉的影响:体外研究

阅读:7
作者:Yufeng Zhou, Wei Chun Daniel Lim

Abstract

High-intensity focused ultrasound (HIFU) has been used to ablate solid tumors and cancers. Because of the hypervascular structure of the tumor and circulating blood inside it, the interaction between the HIFU burst and vessel is a critical issue in the clinical environment. Influences on lesion production and the potential of vessel rupture were investigated in this study for the efficiency and safety of clinical ablation. An extracted porcine artery was embedded in a transparent polyacrylamide gel phantom, with bovine serum albumin (BSA) as an indicator of the thermal lesion, and degassed water was driven through the artery sample. The HIFU focus was aligned to the anterior wall, middle of the artery, and posterior wall. After HIFU ablation, the produced lesion was photographically recorded, and then its size was quantified and compared with that in the gel phantom without artery. In addition, the bubble dynamics (i.e., generation, expansion, motion, and shrinkage of bubbles and their interaction with the artery) were captured using high-speed imaging. It was found that the presence of the artery resulted in a decrease in lesion size in both the axial and lateral directions. The characteristics of the lesion are dependent on the focus alignment. Acoustic and hydrodynamic cavitation play important roles in lesion production and interaction with the artery. Both thermal and mechanical effects were found on the surface of the artery wall after HIFU ablation. However, no vessel rupture was found in this ex vivo study.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。