Whole Blood Titanium Concentration after Limb Salvage Surgery with Three-Dimensional-Printed Ti6Al4V Implants

三维打印 Ti6Al4V 植入物肢体挽救手术后全血钛浓度

阅读:7
作者:Jong Woong Park, Se Hoon Jung, Jung Hwan Yang, June Hyuk Kim, Sung Eun Oh, Hyun Guy Kang

Background

Three-dimensional (3D)-printed customized implants can be fabricated and utilized for all bones with massive bone defects. The main safety issues with 3D-printed implants made of Ti6Al4V alloy are related to the release of metal debris and residual powder. In this study, we investigated the perioperative titanium concentrations in whole blood and peri-implant fluid samples of patients who underwent limb salvage surgery with a 3D-printed Ti6Al4V implant.

Conclusions

Whole blood titanium concentrations were higher after surgery using 3D-printed implants than after that using conventional orthopedic implants, but markedly lower than in patients with implant failure. None of the patients developed serious clinical adverse effects during follow-up.

Methods

Nineteen patients who underwent limb salvage surgery with 3D-printed Ti6Al4V implants were divided into two groups: the serial samples group and the follow-up group. To observe metal distribution and clearance in the body, serial samples of blood and peri-implant fluid from the surgical drain were prospectively collected for five patients in the serial samples group. For the remaining 14 patients who were followed up for more than a year, blood samples were collected only once.

Results

In the serial samples group, the mean baseline titanium concentration was 0.78 µg/L (range, 0.1-2.2 µg/L): 3 patients showed peak concentration before the third postoperative month, while 2 patients still showed an increasing pattern at this point. Total titanium mass in the surgical drain showed a wash-out phenomenon in a week, with a significant uniform decrease (p = 0.04). In 14 patients in the follow-up group, the mean titanium concentration in the whole blood was 10.8 µg/L (range, 0.3-36.6 µg/L). For the 14 patients with a long-term follow-up, the aluminum and vanadium concentrations were all negligible. Conclusions: Whole blood titanium concentrations were higher after surgery using 3D-printed implants than after that using conventional orthopedic implants, but markedly lower than in patients with implant failure. None of the patients developed serious clinical adverse effects during follow-up.