Examination of an absolute quantity of less than a hundred nanograms of proteins by amino acid analysis

通过氨基酸分析检查绝对量少于一百纳克的蛋白质

阅读:6
作者:Akiko Masuda, Naoshi Dohmae

Abstract

We developed an ultra-sensitive method of amino acid analysis (AAA) for the absolute quantification of less than 100 ng of proteins, in solution or on polyvinylidene difluoride (PVDF) membranes using an oxygen-free chamber for protein hydrolysis. We used a pre-label method with 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate for fluorescence detection, ion-pair chromatography with a reversed-phase column, and an ultra-high-pressure high-performance liquid chromatography. We optimized both handling- and instrument-dependent factors for accurate quantification and showed that the least amount of proteins to quantify was determined by handling accuracy rather than instrumental limit for quantification which was 0.6 fmol/amino acid. As a new evaluation method for the handling accuracy, we adopted the protein identification by the obtained amino acid compositions by AAA and the Swiss-Prot database search without the restriction of species. As a result, the least amount of starting material for AAA was 16 ng (0.24 pmol) for a solution of bovine serum albumin (BSA), 33 ng (0.50 pmol) for BSA on a PVDF membrane, and 44 ng (0.15 pmol) for thyroglobulin on a PVDF membrane. These results demonstrate that the ultra-sensitive AAA developed in this study is feasible for absolute quantification of biological significant protein.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。