Conclusions
The detection of empty capsids is unexpected and the process that leads to them is unknown. The average charge on the empty capsids is significantly lower than expected from the charge residue model, which probably indicates that the empty capsids have contracted in the gas phase. The scaffolding protein presumably limits the contraction of the procapsids. This work shows that electrospray CDMS can provide valuable information for masses greater than 20 MDa.
Methods
P22 procapsids containing a native cargo of scaffolding protein were assembled in E. coli and purified via differential centrifugation. Electrospray CDMS was used to measure their mass distribution.
Results
The procapsid peak was centered at 23.60 MDa, which indicates that they contain an average of ~112 scaffolding proteins. The distribution is relatively narrow, less than 31 scaffolding proteins wide. In addition, a peak at 19.84 MDa with a relative abundance of ~15% is attributed to empty capsids. Despite having the same sizes in solution, the empty capsid and the procapsid have significantly different average charges. Conclusions: The detection of empty capsids is unexpected and the process that leads to them is unknown. The average charge on the empty capsids is significantly lower than expected from the charge residue model, which probably indicates that the empty capsids have contracted in the gas phase. The scaffolding protein presumably limits the contraction of the procapsids. This work shows that electrospray CDMS can provide valuable information for masses greater than 20 MDa.