Field evaluation of onsite near real-time monitors for surface contamination by 5-fluorouracil

5-氟尿嘧啶表面污染现场近实时监测器的现场评估

阅读:6
作者:Jerome P Smith, Deborah Sammons, Shirley Robertson, Edward Krieg, John Snawder

Conclusion

The data indicate that there are many areas in hospitals that are contaminated with 5-fluorouracil and the monitors will be useful in identifying this contamination.

Methods

The 5-fluorouracil monitor was evaluated in areas where oncology drugs were prepared and administered to patients at five different hospitals. The levels of contamination measured with the monitors were compared to levels measured with liquid chromatography-tandem mass spectrometry.

Results

The 5-fluorouracil values measured with the liquid chromatography-tandem mass spectrometry ranged from 0 to over 200,000 ng/100 cm2. Measurements by the 5-fluorouracil monitors in the range 10-100 ng/100 cm2 correlated with the liquid chromatography-tandem mass spectrometry. Receiver operating characteristic curves developed for the data indicated that a positive limit of 22 ng/100 cm2 would give an acceptable level of false-positives while retaining most true-positive samples. If the liquid chromatography-tandem mass spectrometry measured greater than 100 ng/100 cm2, then the monitors also measured levels greater than 100 ng/100 cm2 for the majority of samples.