The Effects of Gamma Irradiation on Chemical Biomarker Recovery from Mixed Chemical/Biological Threat Exposure Specimens

伽玛射线照射对混合化学/生物威胁暴露样本中化学生物标志物恢复的影响

阅读:2
作者:Samantha L Isenberg, Melissa D Carter, Jonathan L Moon, Sarah Laughlin, Marla Petway, Mike A Mojica, Julia E Rood, Alexis K Gursky, Cody I Sheppard, Dennis A Bagarozzi, James L Pirkle, Rudolph C Johnson

Background

Irradiative sterilization of clinical specimens prior to chemical laboratory testing provides a way to not only sterilize pathogens and ensure laboratorian safety but also preserve sample volume and maintain compatibility with quantitative chemical diagnostic protocols. Since the compatibility of clinical biomarkers with gamma irradiation is not well characterized, a subset of diagnostic biomarkers ranging in molecular size, concentration, and clinical matrix was analyzed to determine recovery following gamma irradiation.

Conclusions

The results from this pilot study support the implementation of an irradiative sterilization protocol for possible mixed-exposure samples containing both chemical and biological threat agents (mixed CBTs). Furthermore, irradiative sterilization significantly reduces a laboratorian's risk of infection from exposure to an infectious agent without compromising chemical diagnostic testing integrity, particularly for diagnostic assays in which the chemical analyte has been shown to be fully conserved following a 5 Mrad irradiative dose.

Methods

Sample irradiation of previously characterized quality control materials (QCs) at 5 Mrad was carried out at the Gamma Cell Irradiation Facility at the Centers for Disease Control and Prevention (CDC) in Atlanta, GA. Following irradiation, the QCs were analyzed alongside non-irradiated QCs to determine analyte recovery between dosed and control samples.

Results

Biomarkers for exposure to abrin, ricin, and organophosphorus nerve agents (OPNAs) were analyzed for their stability following gamma irradiation. The diagnostic biomarkers included adducts to butyrylcholinesterase, abrine, and ricinine, respectively, and were recovered at over 90% of their initial concentration. Conclusions: The results from this pilot study support the implementation of an irradiative sterilization protocol for possible mixed-exposure samples containing both chemical and biological threat agents (mixed CBTs). Furthermore, irradiative sterilization significantly reduces a laboratorian's risk of infection from exposure to an infectious agent without compromising chemical diagnostic testing integrity, particularly for diagnostic assays in which the chemical analyte has been shown to be fully conserved following a 5 Mrad irradiative dose.