Towards bedside washing of stored red blood cells: a prototype of a simple apparatus based on microscale sedimentation in normal gravity

床边清洗储存红细胞:基于常重力微尺度沉降的简单装置原型

阅读:11
作者:G Khanal, R A Huynh, K Torabian, H Xia, E Vörös, S S Shevkoplyas

Conclusions

This proof-of-concept study demonstrates the feasibility of washing stored RBCs using a simple, disposable system with efficiency comparable to that of conventional centrifugation, and thus represents a significant first step towards enabling low-cost washing of stored blood at bedside.

Methods

Stored RBCs were diluted to 10% haematocrit (Hct) with normal saline, matching the conventional washing procedure. The dilute RBC suspensions were passed through a column of coiled tubing to allow RBC sedimentation in normal gravity, thus separating them from the washing solution. Washed RBCs were collected using bifurcations located along the tubing. Washing efficiency was quantified by measuring Hct, morphology, deformability, free haemoglobin and total-free protein.

Results

The gravity-driven washing system operating at 0·5 ml/min produced washed RBCs with final Hct of 36·7 ± 3·4% (32·3-41·2%, n = 10) and waste Hct of 3·4 ± 0·7% (2·4-4·3%, n = 10), while removing 80% of free haemoglobin and 90% of total-free protein. Washing improved the ability of stored RBCs to perfuse an artificial microvascular network by 20%. The efficiency of washing performed using the gravity-driven system was not significantly different than that of conventional centrifugation. Conclusions: This proof-of-concept study demonstrates the feasibility of washing stored RBCs using a simple, disposable system with efficiency comparable to that of conventional centrifugation, and thus represents a significant first step towards enabling low-cost washing of stored blood at bedside.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。