Analytical performance of publicly dispensed glucometers in primary health care in a southern Brazilian city

巴西南部城市初级卫生保健中公用血糖仪的分析性能

阅读:18
作者:Isabelle L Silva, Flávia Martinello

Aims

This study aimed to assess the use of glucometers by patients and the analytical performance of glucometers provided by the primary care services.

Conclusions

Overall, our findings showed that patients had a good understanding of glucometer usage and suggested that some glucometers should be replaced, as they sometimes failed to meet even the manufacturer's acceptable variation limits, and/or did not meet QSBV.

Methods

The analytical performance of 48 glucometers Accu-Chek® Active, was assessed through quintuplicate analyses of one Roche and one PNCQ (National Quality Control Program) control sample at different concentrations; 31 were also evaluated by a single proficiency testing sample. The evaluation metrics included imprecision, bias, and total error and were measured according to quality specifications based on biological variation (QSBV). Glucometer users answered a questionnaire regarding their experience.

Results

Among the 48 glucometers evaluated with internal control samples, 17 met precision criteria at both control levels according to QSBV, while 24 met the criteria at only one control level. Of the 31 glucometers further evaluated through proficiency test, 11 met accuracy criteria according to QSBV, and only one device showed an unacceptable result. Out of these 31, only 15 demonstrated a total error within the acceptable maximum limits based on QSBV. Conclusions: Overall, our findings showed that patients had a good understanding of glucometer usage and suggested that some glucometers should be replaced, as they sometimes failed to meet even the manufacturer's acceptable variation limits, and/or did not meet QSBV.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。