Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers

新生儿血红蛋白影响GEM Premier 4000血气分析仪全血胆红素测量的准确性

阅读:5
作者:Yun Huang, Robert Dean, Yvonne Dubbelman, Anne Vincent, Faiza Khurshid

Conclusions

Neonatal whole blood bilirubin measurement on blood gas analyzers may be affected by neonatal hemoglobin. The method should be validated using neonatal whole blood samples or samples with a similar matrix before the analyzers are implemented into neonatal hyperbilirubinemia management.

Methods

Bilirubin in adult and umbilical cord whole blood samples, spiked with an unconjugated bilirubin standard, and non-spiked adult plasma samples was measured on a blood gas analyzer (GEM 4000) and a Core Laboratory Chemistry analyzer (Architect c16000) respectively. We also investigated the linear regression for neonatal and adult hemoglobin measured on the blood gas analyzer and the Core Laboratory hematology analyzer (Alinity h-Series).

Results

Plasma bilirubin measured on the blood gas analyzer and the chemistry analyzer was statistically identical. Adult whole blood bilirubin showed slightly increased proportional bias. When umbilical cord whole blood samples were used, the Deming regression showed GEM bilirubin =1.233(Architect) (95% CI 1.199 ~ 1.266)-44.43 ​μmol/L (95% CI -53.6 ​~ ​-35.2). The regression was significantly different from that in plasma (p ​< ​0.001) or adult whole blood (p ​< ​0.001) samples. 36.1% neonatal samples with bilirubin levels >50 ​μmol/L showed that the bias% was above laboratory standards. In addition, the regression of neonatal hemoglobin measurement between the GEM and the Alinity was significantly different from adult hemoglobin (p ​< ​0.01). Conclusions: Neonatal whole blood bilirubin measurement on blood gas analyzers may be affected by neonatal hemoglobin. The method should be validated using neonatal whole blood samples or samples with a similar matrix before the analyzers are implemented into neonatal hyperbilirubinemia management.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。