Placental perfusion in uterine ischemia model as evaluated by dynamic contrast enhanced MRI

动态增强磁共振成像评价子宫缺血模型中的胎盘灌注

阅读:10
作者:Alexander Drobyshevsky, P V Prasad

Background

To validate DCE MRI method of placental perfusion estimation and to demonstrate application of the method in a rabbit model of fetal antenatal hypoxia-ischemia.

Conclusion

Underestimation of true perfusion in placenta by steepest slope DCE MRI is significant and the error appears to be systematic.

Methods

Placental perfusion was estimated by dynamic contrast imaging with bolus injection of Gd-DTPA in 3 Tesla GE magnet in a rabbit model of placental ischemia-reperfusion in rabbit dams at embryonic day 25 gestation age. Placental perfusion was measured using steepest slope method on DCE MRI before and after intermittent 40 min uterine ischemia. Antioxidants (n = 2 dams, 9 placentas imaged) or vehicle (n = 5 dams, 23 placenta imaged) were given systemically in a separate group of dams during reperfusion-reoxygenation. Placental perfusion was also measured in two dams from the antioxidant group (10 placentas) and two dams from the control group (12 placentas) by fluorescent microspheres method.

Results

While placental perfusion estimates between fluorescent microspheres and DCE MRI were significantly correlated (R(2) = 0.85; P < 0.01), there was approximately 33% systematic underestimation by the latter technique. DCE MRI showed a significant decrease in maternal placental perfusion in reperfusion-reoxygenation phase in the saline, 0.44 ± 0.06 mL/min/g (P = 0.012, t-test), but not in the antioxidant group, 0.62 ± 0.06 mL/min/g, relative to pre-occlusion values (0.77 ± 0.07 and 0.84 ± 0.12 mL/min/g, correspondingly).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。