Filling in the gaps with non-standard body fluids

用非标准体液填补空白

阅读:6
作者:Sheng-Ying Lo, Nabiha H Saifee, Brook O Mason, Dina N Greene

Conclusions

Chemistry analytes measured on the AU680 demonstrate acceptable analytical performance for non-standard fluids. Common endogenous interference from lipemia, icterus, and hemolysis (LIH) are observed and flagging rules based on LIH indices were developed to help improve the clinical interpretation of results.

Methods

Historical measurements in non-standard body fluids run on a Beckman Coulter DxC800 were used to optimize population-specific clinical reportable ranges for albumin, amylase, creatinine, glucose, lactate dehydrogenase, lipase, total bilirubin, total cholesterol, total protein, triglyceride and urea nitrogen run on the Beckman Coulter AU680. For these 11 analytes, interference studies were performed by spiking hemolysate, bilirubin, or Intralipid® into abnormal serous fluids. Precision, accuracy, linearity, and stability of total bilirubin in non-standard fluids was evaluated on the Beckman Coulter AU680 analyzer.

Results

The historical non-standard fluid results indicated that in order to report a numeric result, 4 assays required no dilution, 5 assays required onboard dilutions and 2 assays required both onboard and manual dilutions. The AU680 total bilirubin assay is suitable for clinical testing of non-standard fluids. Interference studies revealed that of the 11 total AU680 analyte measurements on non-standard fluids, lipemia affected 1, icterus affected 3, and hemolysis affected 5. Conclusions: Chemistry analytes measured on the AU680 demonstrate acceptable analytical performance for non-standard fluids. Common endogenous interference from lipemia, icterus, and hemolysis (LIH) are observed and flagging rules based on LIH indices were developed to help improve the clinical interpretation of results.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。