Vaginal degeneration following implantation of synthetic mesh with increased stiffness

植入硬度增加的合成网片后阴道退化

阅读:9
作者:R Liang, S Abramowitch, K Knight, S Palcsey, A Nolfi, A Feola, S Stein, P A Moalli

Conclusion

Mesh implantation with the stiffer mesh Gynemesh PS induced a maladaptive remodelling response consistent with vaginal degeneration.

Methods

Following Institutional Animal Care Use Committee approval, 50 rhesus macaques were implanted with Gynemesh PS (n = 12), UltraPro with its blue line perpendicular to the longitudinal axis of vagina (n = 10), UltraPro with its blue line parallel to the longitudinal axis of vagina (n = 8) or SmartMesh (n = 8) via sacrocolpopexy following hysterectomy. Sham-operated animals (n = 12) served as controls. Main outcome measures: The mesh-vagina complex was removed after 12 weeks and analysed for histomorphology, in situ cell apoptosis, total collagen, elastin, glycosaminoglycan content and total collagenase activity. Appropriate statistics and correlation analyses were performed accordingly.

Objective

To compare the impact of the prototype prolapse mesh Gynemesh PS with that of two new-generation lower stiffness meshes, UltraPro and SmartMesh, on vaginal morphology and structural composition. Design: A mechanistic study employing a nonhuman primate model. Setting: Magee-Womens Research Institute at the University of Pittsburgh. Population: Parous rhesus macaques, with similar age, weight, parity and Pelvic Organ Prolapse-Questionnaire scores.

Results

Relative to sham and the two lower stiffness meshes, Gynemesh PS had the greatest negative impact on vaginal histomorphology and composition. Compared with sham, implantation with Gynemesh PS caused substantial thinning of the smooth muscle layer (1557 ± 499 μm versus 866 ± 210 μm, P = 0.02), increased apoptosis particularly in the area of the mesh fibres (P = 0.01), decreased collagen and elastin content (20%, P = 0.03 and 43%, P = 0.02, respectively) and increased total collagenase activity (135%, P = 0.01). Glycosaminoglycan, a marker of tissue injury, was highest with Gynemesh PS compared with sham and other meshes (P = 0.01).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。