Neonatal sepsis is still a leading cause of death among newborns. Therefore a protein-microarray for point-of-care testing that simultaneously quantifies the sepsis associated serum proteins IL-6, IL-8, IL-10, TNF alpha, S-100, PCT, E-Selectin, CRP and Neopterin has been developed. The chip works with only a 4 μL patient serum sample and hence minimizes excessive blood withdrawal from newborns. The 4 μL patient samples are diluted with 36 μL assay buffer and distributed to four slides for repetitive measurements. Streptavidin coated magnetic particles that act as distinct stirring detection components are added, not only to stir the sample, but also to detect antibody antigen binding events. We demonstrate that the test is complete within 2.5 h using a single step assay. S-100 conjugated to BSA is spotted in increasing concentrations to create an internal calibration. The presented low volume protein-chip fulfills the requirements of point-of-care testing for accurate and repeatable (CV < 14%) quantification of serum proteins for the diagnosis of neonatal sepsis.
Miniaturized protein microarray with internal calibration as point-of-care device for diagnosis of neonatal sepsis.
具有内部校准功能的微型蛋白质微阵列,可作为新生儿败血症诊断的即时检测设备
阅读:5
作者:Buchegger Patricia, Sauer Ursula, Toth-Székély Hedvig, Preininger Claudia
| 期刊: | Sensors | 影响因子: | 3.500 |
| 时间: | 2012 | 起止号: | 2012;12(2):1494-508 |
| doi: | 10.3390/s120201494 | 研究方向: | 炎症/感染 |
| 疾病类型: | 败血症 | ||