INTRODUCTION: Point-of-care (POC) CD4 testing can improve access to treatment by enabling decentralization and reducing patient loss-to-follow-up. As new POC CD4 technologies become available, their performance should be assessed before widespread deployment. This study reports the findings of five independent evaluations of the PointCare NOW CD4 system. MATERIALS/METHODS: Evaluations were conducted in Southern Africa (Mozambique, South Africa) and North America (Canada, USA). 492 blood samples (55 from HIV-negative blood donors and 437 from HIV-infected patients, including 20 children aged between 12 and 59 months) were tested with both the PointCare NOW and reference flow cytometry instruments. Assessment of bias, precision and levels of clinical misclassification for absolute and percent CD4 count was conducted. RESULTS: PointCare NOW significantly overestimated CD4 absolute counts with a mean relative bias of +35.0%. Bias was greater in samples with CD4 counts below ⤠350 cells/µl (+51.3%) than in the CD4 >350 cells/µl stratum (15.1%). Bias in CD4% had a similar trend with an overall relative mean bias of +25.6% and a larger bias for low CD4 stratum (+40.2%) than the higher CD4 stratum (+5.8%). Relative bias for CD4% in children was -6.8%. In terms of repeatability, PointCare NOW had a coefficient of variation of 11%. Using a threshold of 350 cells/µl, only 47% of patients who qualified for antiretroviral therapy with reference CD4 testing, would have been eligible for treatment with PointCare NOW test results. This was 39% using a 200 cells/µl threshold. Agreement with infant samples was higher, with 90% qualifying at a 25% eligibility threshold. CONCLUSION: The performance of the PointCare NOW instrument for absolute and percent CD4 enumeration was inadequate for HIV clinical management in adults. In children, the small sample size was not large enough to draw a conclusion. This study also highlights the importance of independent evaluation of new diagnostic technology platforms before deployment.
Performance of the PointCare NOW system for CD4 counting in HIV patients based on five independent evaluations.
PointCare NOW 系统在 HIV 患者 CD4 计数中的性能基于五项独立评估
阅读:4
作者:Bergeron Michèle, Daneau Géraldine, Ding Tao, Sitoe Nadia E, Westerman Larry E, Stokx Jocelijn, Jani Ilesh V, Coetzee Lindi M, Scott Lesley, De Weggheleire Anja, Boel Luc, Stevens Wendy S, Glencross Deborah K, Peter Trevor F
| 期刊: | PLoS One | 影响因子: | 2.600 |
| 时间: | 2012 | 起止号: | 2012;7(8):e41166 |
| doi: | 10.1371/journal.pone.0041166 | 研究方向: | 其它 |
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。