BACKGROUND: The hemocompatibility of dialyzers for extracorporeal kidney replacement therapy (KRT) is of importance to minimize harmful reactions between blood constituents and the membrane. We investigated in these exploratory studies the hemocompatibility profile of several types of polysulfone dialyzers. METHODS: Hemocompatibility of various high-flux polysulfone dialyzers were compared in two consecutive, prospective, randomized, crossover studies, each including 24 adult patients being at least 3 months on hemodialysis (HD) or on-line hemodiafiltration (HDF). These dialyzers, differing in membrane type, fiber geometry, sterilization method, and production technology, were each applied for 1 week in HD or HDF. Hemocompatibility was assessed through markers of complement activation, cell activation, coagulation, contact activation, and immunologic reactions. RESULTS: The patients in the two studies were on average 67±11 and 68±11 years old, 75% and 67% were male, and were on KRT for 5.4±5.0 and 4.4±3.6 years. The complement factors C3a and C5a increased early and transiently during treatment, less so with HDF than with HD, and with dialyzers combining wider inner fiber diameter (210 versus 185 µm) and advanced membrane type (Helixone plus versus Helixone). sC5b-9 increased in all study phases, reaching its highest level after 60 minutes, with lower values over the entire treatment (area under the curve) for HDF than HD, and for wider inner fiber diameter and advanced membrane type. Leukocytes decreased in the first 10 minutes, without significant differences between dialyzers. PMN elastase increased in the first hour, more so with HD than HDF. Thrombocytes decreased slightly in the first 30 minutes, with differences only between HDF and HD mode. IL-8 decreased from pre- to postdialysis, particularly on HDF. No differences were observed for kallikrein, IgE, and hsCRP. CONCLUSIONS: In these explorative studies we found indications to a comparable hemocompatibility profile of the investigated dialyzers. We observed distinctions in compounds between HDF and HD and for some dialyzer and membrane characteristics.
Hemocompatibility of Polysulfone Hemodialyzers - Exploratory Studies on Impact of Treatment Modality and Dialyzer Characteristics.
聚砜血液透析器的血液相容性——治疗方式和透析器特性影响的探索性研究
阅读:4
作者:Wagner Stephan, Zschätzsch Sebastian, Erlenkoetter Ansgar, Rauber Lena, Stauss-Grabo Manuela, Gauly Adelheid
| 期刊: | Kidney360 | 影响因子: | 3.000 |
| 时间: | 2020 | 起止号: | 2020 Jan 9; 1(1):25-35 |
| doi: | 10.34067/KID.0000342019 | 研究方向: | 其它 |
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。