Understanding the aliya pulsed electric field dose-response relationship: Implications for ablation size, thermal load, and immune response in an orthotopic murine breast cancer model.

了解 aliya 脉冲电场剂量反应关系:对原位小鼠乳腺癌模型中消融面积、热负荷和免疫反应的影响

阅读:4
作者:Nafie Ebtesam H O, Pastori Chiara, Acharya Rupsa, Kaviani Rosa, Hunter David, Shalaby Waleed, Krimsky William S, Neal Robert E 2nd
Aliya®Pulsed Electric Field (PEF) technology is an emerging strategy in the field of cancer treatment, offering a novel approach to ablation therapy that does not rely on thermal mechanisms. By employing a multi-stage experimental setup, including potato tuber, porcine liver, and murine breast cancer models, we explored the dose-response relationship on ablation and immune modulation by varying the pulse packets delivered from 20 to 100. The biologic response observed with 60 packets represented a minimum effective dose yielding reproducible ablation parameters, immune response, and efficacy which could be augmented with immune checkpoint blockade. This pre-clinical analysis provides a first step toward understanding the therapeutic index for PEF technology beyond ablation, a consideration that will require robust clinical validation in well-designed prospective studies.