How Geometry Affects Sensitivity of a Differential Transformer for Contactless Characterization of Liquids.

几何形状如何影响差动变压器在液体非接触式特性分析中的灵敏度

阅读:4
作者:Berger Marc, Zygmanowski Anne, Zimmermann Stefan
The electrical and dielectric properties of liquids can be used for sensing. Specific applications, e.g., the continuous in-line monitoring of blood conductivity as a measure of the sodium concentration during dialysis treatment, require contactless measuring methods to avoid any contamination of the medium. The differential transformer is one promising approach for such applications, since its principle is based on a contactless, magnetically induced conductivity measurement. The objective of this work is to investigate the impact of the geometric parameters of the sample or medium under test on the sensitivity and the noise of the differential transformer to derive design rules for an optimized setup. By fundamental investigations, an equation for the field penetration depth of a differential transformer is derived. Furthermore, it is found that increasing height and radius of the medium is accompanied by an enhancement in sensitivity and precision.