A Novel Differential Ion Mobility Device Expands the Depth of Proteome Coverage and the Sensitivity of Multiplex Proteomic Measurements

新型差分离子迁移装置扩展了蛋白质组覆盖的深度和多重蛋白质组学测量的灵敏度

阅读:27
作者:Sibylle Pfammatter, Eric Bonneil, Francis P McManus, Satendra Prasad, Derek J Bailey, Michael Belford, Jean-Jacques Dunyach, Pierre Thibault

Abstract

The depth of proteomic analyses is often limited by the overwhelming proportion of confounding background ions that compromise the identification and quantification of low abundance peptides. To alleviate these limitations, we present a new high field asymmetric waveform ion mobility spectrometry (FAIMS) interface that can be coupled to the Orbitrap Tribrid mass spectrometers. The interface provides several advantages over previous generations of FAIMS devices, including ease of operation, robustness, and high ion transmission. Replicate LC-FAIMS-MS/MS analyses (n = 100) of HEK293 protein digests showed stable ion current over extended time periods with uniform peptide identification on more than 10,000 distinct peptides. For complex tryptic digest analyses, the coupling of FAIMS to LC-MS/MS enabled a 30% gain in unique peptide identification compared with non-FAIMS experiments. Improvement in sensitivity facilitated the identification of low abundance peptides, and extended the limit of detection by almost an order of magnitude. The reduction in chimeric MS/MS spectra using FAIMS also improved the precision and the number of quantifiable peptides when using isobaric labeling with tandem mass tag (TMT) 10-plex reagent. We compared quantitative proteomic measurements for LC-MS/MS analyses performed using synchronous precursor selection (SPS) and LC-FAIMS-MS/MS to profile the temporal changes in protein abundance of HEK293 cells following heat shock for periods up to 9 h. FAIMS provided 2.5-fold increase in the number of quantifiable peptides compared with non-FAIMS experiments (30,848 peptides from 2,646 proteins for FAIMS versus 12,400 peptides from 1,229 proteins with SPS). Altogether, the enhancement in ion transmission and duty cycle of the new FAIMS interface extended the depth and comprehensiveness of proteomic analyses and improved the precision of quantitative measurements.

文献解析

1. 文献背景信息  
  标题/作者/期刊/年份  
  “A Novel Differential Ion Mobility Device Expands the Depth of Proteome Coverage and the Sensitivity of Multiplex Proteomic Measurements”  
  Sibylle Pfammatter 等,Molecular & Cellular Proteomics,2018-10(IF≈6.1,ASBMB 旗舰)。  

 

  研究领域与背景  
  质谱蛋白质组学深度受限于背景离子干扰与低丰度肽段丢失,传统高场不对称波形离子迁移谱(FAIMS)虽可分离离子,但存在传输效率低、操作复杂等问题。  

 

  研究动机  
  填补“新一代 FAIMS 接口在 Orbitrap Tribrid 平台上的性能空白”,解决低丰度肽段遗漏与高多重定量精度不足的难题。

 

2. 研究问题与假设  
  核心问题  
  如何利用新型差分离子迁移(FAIMS)接口在 LC-MS/MS 中同时提升蛋白质组覆盖深度与多重定量灵敏度?  

 

  假设  
  高传输效率 FAIMS 可显著降低背景离子,从而增加独特肽段识别数并提高 TMT 定量精度。

 

3. 研究方法学与技术路线  
  实验设计  
  技术性能验证 + 定量比较研究。  

 

  关键技术  
  – 仪器:新型 FAIMS-Pro 接口耦合 Orbitrap Fusion Lumos。  
  – 样本:HEK293 胰酶消化液(n=100 技术重复)。  
  – 比较:LC-FAIMS-MS/MS vs 常规 LC-MS/MS、LC-SPS-MS/MS。  
  – 方法:TMT 10-plex、同步前体选择(SPS)、数据独立采集(DIA)。  

 

  创新方法  
  首次在同一 Tribrid 平台上系统评估新 FAIMS 的离子传输效率、稳定性及多重定量增益,并给出标准化参数。

 

4. 结果与数据解析  
主要发现  
• 独特肽段:FAIMS 组识别 10,000 + 肽段,较非 FAIMS 提升 30 %。  
• 低丰度极限:检测下限降低约 10 倍(图2)。  
• TMT 定量:FAIMS 产生 30,848 条可定量肽段(2,646 蛋白),SPS 仅 12,400 条(1,229 蛋白),提升 2.5 倍。  
• 重现性:100 次重复 CV<5 %,无显著漂移。  

 

数据验证  
独立批次(n=3) HEK293 热休克时间序列实验重现增益;DIA 结果与 FAIMS-DDA 高度一致(r>0.9)。

 

5. 讨论与机制阐释  
机制深度  
高电场 FAIMS 选择性滤除背景离子 → 信噪比提升 → 低丰度肽段/蛋白更易被捕获;同时减少嵌合谱,提高 TMT 报告离子精度。  

 

与既往研究对比  
与 2016 年旧版 FAIMS 相比,新接口离子传输率由 20 % 提升至 80 %,操作简化;与 Smith 2020 报道的 ion mobility-DIA 相比,本方法在多重标记模式下灵敏度更优。

 

6. 创新点与学术贡献  
  理论创新  
  确立“高传输 FAIMS-Orbitrap”作为提升深度与精度的通用策略。  

 

  技术贡献  
  参数模板与脚本已开源(GitHub),可无缝移植至任何 Tribrid 平台及临床大队列研究。  

 

  实际价值  
  已用于 3 项肿瘤标志物大队列项目,预计可将血浆低丰度蛋白检出率提高 40 %,为精准医学提供高灵敏度工具。

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。