In this study, we investigated a label-free time efficient biosensor to recognize growth factors (GF) in real time, which are of gran interesting in the regulation of cell division and tissue proliferation. The sensor is based on a system of shear horizontal surface acoustic wave (SH-SAW) immunosensor combined with a microfluidic chip, which detects GF samples in a dynamic mode. In order to prove this method, to our knowledge not previously used for this type of compounds, two different GFs were tested by two immunoreactions: neurotrophin-3 and fibroblast growth factor-2 using its polyclonal antibodies. GF detection was conducted via an enhanced sequential workflow to improve total test time of the immunoassay, which shows that this type of biosensor is a very promising method for ultra-fast recognition of these biomolecules due to its great advantages: portability, simplicity of use, reusability, low cost, and detection within a relatively short period of time. Finally, the biosensor is able to detect FGF-2 growth factor in a concentration wide range, from 1-25 µg/mL, for a total test time of ~15 min with a LOD of 130 ng/mL.
Novel SH-SAW Biosensors for Ultra-Fast Recognition of Growth Factors.
用于超快速识别生长因子的新型SH-SAW生物传感器
阅读:5
作者:Matatagui Daniel, Bastida Ãgatha, Horrillo M Carmen
| 期刊: | Biosensors-Basel | 影响因子: | 5.600 |
| 时间: | 2021 | 起止号: | 2021 Dec 30; 12(1):17 |
| doi: | 10.3390/bios12010017 | 研究方向: | 其它 |