Analysis of Singlet Oxygen Luminescence Generated By Protoporphyrin IX.

对原卟啉IX产生的单线态氧发光进行分析

阅读:4
作者:Vikas Vikas, Yang Weibing, Wilson Brian C, Zhu Timothy C, Hadfield Robert H
The effectiveness of photodynamic therapy (PDT) for cancer treatment relies on the generation of cytotoxic singlet oxygen ((1)O(2)) in type II PDT. Hence, monitoring of (1)O(2) generation during PDT enables optimal treatment delivery to the tumor target with reduced off-target effects. Direct (1)O(2) observation by measuring its luminescence at 1270 nm remains challenging due to the very weak signal. This study presents (1)O(2) luminescence measurements using a time-resolved singlet oxygen luminescence detection system (TSOLD) applied to protoporphyrin IX (PpIX) in different solvents (ethanol and acetone) and biological media (bovine serum albumin and agarose-based solid phantom). The compact experimental setup includes a nanosecond diode laser with a function generator, a cuvette with photosensitizer solution, optical filtering and mirrors, an InGaAs single-photon avalanche diode detector, and time-tagger electronics. Increasing the concentration of PpIX in these media from 1 to 10 µg/g resulted in a 3-5 × increase in the (1)O(2) luminescence signal. Furthermore, increasing light scattering in the sample using Intralipid from 0.1 to 1% led to a decrease in the (1)O(2) luminescence signal and lifetime. These results confirm the marked effect of the microenvironment on the (1)O(2) signal and, hence, on the photodynamic efficacy.