PURPOSE: The aim of this study is to correlate the intraocular pressure (IOP) change with the acoustic impedance of the cornea, in order to propose a noncontact and noninvasive method for IOP monitoring. METHODS AND MATERIALS: A highly focused transducer (frequency 47-MHz; bandwidth 62%) was made to measure the echo from the anterior and posterior surfaces of intact porcine eyes, respectively. A multilayered transmission and reflection model was used to calculate the acoustic impedance. The linear relationship between acoustic impedance and intraocular pressure was analyzed by statistical method. RESULT: During pressure elevation from 10âmm Hg to 50âmm Hg, the mean acoustic impedance of the posterior cornea increased from 1.5393 to 1.5698âMRayl, which showed a strong linear correlation (R = 0.9849; P = 0.0022). Meanwhile, the mean value of the anterior cornea increased from 1.5399 to 1.5519âMRayl, and a less significant correlation was observed (R = 0.7378; P = 0.0025). CONCLUSION: This study revealed a linear correlation between intraocular pressure and acoustic impedance of the cornea, thus demonstrating a potentially important method to noninvasively measure the intraocular pressure in vivo.
Correlation of IOP with Corneal Acoustic Impedance in Porcine Eye Model.
阅读:3
作者:Zhang Jun, Zhang Yi, Li Yang, Chen Ruimin, Shung K Kirk, Richter Grace, Zhou Qifa
| 期刊: | Biomed Research International | 影响因子: | 2.300 |
| 时间: | 2017 | 起止号: | 2017;2017:2959717 |
| doi: | 10.1155/2017/2959717 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。