This article presents a new working electrode based on a bismuth-plated, gold-based microelectrode array, which is suitable for determining thallium(I) species using anodic stripping voltammetry (ASV). It allowed a significant increase in the sensitivity as compared to other voltammetric sensors. The main experimental conditions and the instrumental parameters were optimized. A very good proportionality between the Tl(I) peak current and its concentration was evidenced in the range from 5 Ã 10(-10) up to 5 Ã 10(-7) mol L(-1) (R = 0.9989) for 120 s of deposition and from 2 Ã 10(-10) up to 2 Ã 10(-7) mol L(-1) (R = 0.9988) for 180 s. A limit of detection (LOD) of 8 Ã 10(-11) mol L(-1) for a deposition time of 180 s was calculated. The effects of interfering ions on the Tl(I) analytical signal were studied. The proposed method was applied for quantitative Tl(I) detection in water certified reference material TM 25.5 as well as in spiked real water samples, for which satisfactory recovery values between 98.7 and 101.8% were determined.
Anodic Stripping Voltammetric Procedure of Thallium(I) Determination by Means of a Bismuth-Plated Gold-Based Microelectrode Array.
阅读:3
作者:Korolczuk Mieczyslaw, Ochab Mateusz, GÄca Iwona
| 期刊: | Sensors | 影响因子: | 3.500 |
| 时间: | 2024 | 起止号: | 2024 Feb 13; 24(4):1206 |
| doi: | 10.3390/s24041206 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。