Two manganese(III) porphyrins: manganese(III) tetraphenylporphyrin chloride and manganese(III)-tetrakis(3-hydroxyphenyl)porphyrin chloride were tested as ionophores for the construction of new diclofenac-selective electrodes. The electroactive material was incorporated either in PVC or a sol-gel matrix. The effect of different plasticizers and additives (anionic and cationic) on the potentiometric response was studied. The best results were obtained for the PVC membrane plasticized with dioctylphtalate and having sodium tetraphenylborate as a lipophilic anionic additive incorporated. The sensor response was linear in the concentration range 3 Ã 10(-6) - 1 Ã 10(-2) M with a slope of -59.7 mV/dec diclofenac, a detection limit of 1.5 Ã 10(-6) M and very good selectivity coefficients. It was used for the determination of diclofenac in pharmaceutical preparations, by direct potentiometry. The results were compared with those obtained by the HPLC reference method and a good agreement was found between the two methods.
Manganese(III) porphyrin-based potentiometric sensors for diclofenac assay in pharmaceutical preparations.
阅读:3
作者:Vlascici Dana, Pruneanu Stela, Olenic Liliana, Pogacean Florina, Ostafe Vasile, Chiriac Vlad, Pica Elena Maria, Bolundut Liviu Calin, Nica Luminita, Fagadar-Cosma Eugenia
| 期刊: | Sensors | 影响因子: | 3.500 |
| 时间: | 2010 | 起止号: | 2010;10(10):8850-64 |
| doi: | 10.3390/s101008850 | ||