A standard calibration method based on a symmetric resistance network matrix for galvanic logging instruments.

阅读:3
作者:Ji Yongli, Li Zhiqiang, Xia Jigen, Song Jiajia
An increasing number of measurement electrodes have been designed to satisfy the demand for high-resolution detection using galvanic logging technology in complex formations. The forward modeling response analysis of logging tools has important guiding significance in the design of galvanic logging tools. Based on a three-dimensional finite element numerical simulation method, we established a forward model of galvanic multi-electrodes in a complex formation. We also designed a symmetrical resistance network model of the formation with equivalent resistance between two electrodes. A symmetrical resistance network was derived using the balanced bridge method. The asymmetrical admittance matrix was extended to a symmetrical extended admittance matrix to realize a convenient calculation of the equivalent symmetrical resistance network in complex formations. Verification of the microcolumn-focused logging tool, with nine electrodes in a simulated standard well, and an evaluation of the degree of invasion in an actual oil well indicate that this calibration method can improve the measurement accuracy of galvanic logging instruments.