Full-field quantitative phase and polarisation-resolved imaging through an optical fibre bundle.

阅读:3
作者:Gordon George S D, Joseph James, Sawyer Travis, Macfaden Alexander J, Williams Calum, Wilkinson Timothy D, Bohndiek Sarah E
Flexible optical fibres, used in conventional medical endoscopy and industrial inspection, scramble phase and polarisation information, restricting users to amplitude-only imaging. Here, we exploit the near-diagonality of the multi-core fibre (MCF) transmission matrix in a parallelised fibre characterisation architecture, enabling accurate imaging of quantitative phase (error <0.3 rad) and polarisation-resolved (errors <10%) properties. We first demonstrate accurate recovery of optical amplitude and phase in two polarisations through the MCF by measuring and inverting the transmission matrix, and then present a robust Bayesian inference approach to resolving 5 polarimetric properties of samples. Our method produces high-resolution (9.0±2.6μm amplitude, phase; 36.0±10.4μm polarimetric) full-field images at working distances up to 1mm over a field-of-view up to 750×750μm (2) using an MCF with potential for flexible operation. We demonstrate the potential of using quantitative phase for computational image focusing and polarisation-resolved properties in imaging birefringence.