Fast, interleaved, Look-Locker-based T(1) mapping with a variable averaging approach: Towards temperature mapping at low magnetic field.

阅读:4
作者:Fiorito Marco, Yushchenko Maksym, Cicolari Davide, Sarracanie Mathieu, Salameh Najat
Proton resonance frequency shift (PRFS) is currently the gold standard method for magnetic resonance thermometry. However, the linearity between the temperature-dependent phase accumulation and the static magnetic field B(0) confines its use to rather high-field scanners. Applications such as thermal therapies could naturally benefit from lower field MRI settings through leveraging increased accessibility, a lower physical and economical footprint, and further consideration of the technical challenges associated with the integration of heating systems into conventional clinical scanners. T1 -based thermometry has been proposed as an alternative to the gold standard; however, because of longer acquisition times, it has found clinical use solely with adipose tissue where PRFS fails. At low field, the enhanced T1 dispersion, combined with reduced relaxation times, make T1 mapping an appealing candidate. Here, an interleaved Look-Locker-based T1 mapping sequence was proposed for temperature quantification at 0.1 T. A variable averaging scheme was introduced, to maximize the signal-to-noise ratio throughout T1 recovery. In calibrated samples, an average T1 accuracy of 85% ± 4% was achieved in 10 min, compared with the 77% ± 7% obtained using a standard averaging scheme. Temperature maps between 29.0 and 41.7°C were eventually reconstructed, with a precision of 3.0 ± 1.1°C and an accuracy of 1.5 ± 1.0°C. Accounting for longer thermal treatments and less strict temperature constraints, applications such as MR-guided mild hyperthermia treatments at low field could be envisioned.

特别声明

1、本文转载旨在传播信息,不代表本网站观点,亦不对其内容的真实性承担责任。

2、其他媒体、网站或个人若从本网站转载使用,必须保留本网站注明的“来源”,并自行承担包括版权在内的相关法律责任。

3、如作者不希望本文被转载,或需洽谈转载稿费等事宜,请及时与本网站联系。

4、此外,如需投稿,也可通过邮箱info@biocloudy.com与我们取得联系。