BACKGROUND: Although extracorporeal membrane oxygenation (ECMO) has been used to provide temporary support for pediatric patients suffering severe respiratory or cardiac failure since 1970, ECMO systems specifically designed for pediatric patients, particularly for long-term use, remain an unmet clinical need. We sought to develop a new pediatric ECMO system, that is, pediatric pump-lung (PPL), consisting of a unique cylinder oxygenator with an outside-in radial flow path and a centrifugal pump. METHODS: Computational fluid dynamics was used to analyze the blood fluid field for optimized biocompatible and gas exchange performances in terms of flow characteristics, hemolysis, and gas transfer efficiency. Ovine blood was used for in vitro hemolysis and gas transfer testing. RESULTS: Both the computational and experimental data showed that the pressure drop through the PPL's oxygenator is significantly low, even at a flow rate of more than 3.5âL/min. The PPL showed better hemolysis performance than a commercial ECMO circuit consisting of the Quadrox-iD pediatric oxygenator and the Rotaflow pump at a 3.5âL/min flow rate and 250âmm Hg afterload pressure. The oxygen transfer rate of the PPL can reach over 200âmL/min at a flow rate of 3.5âL/min. CONCLUSIONS: The PPL has the potential to provide adequate blood pumping and excellent respiratory support with minimal risk of hemolysis for a wide range of pediatric patients.
Computational fluid dynamics-based design and in vitro characterization of a novel pediatric pump-lung.
阅读:6
作者:Han Dong, Zhang Jiafeng, He Ge, Griffith Bartley P, Wu Zhongjun J
| 期刊: | Artificial Organs | 影响因子: | 2.300 |
| 时间: | 2024 | 起止号: | 2024 Feb;48(2):130-140 |
| doi: | 10.1111/aor.14665 | ||
特别声明
1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。
2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。
3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。
4、投稿及合作请联系:info@biocloudy.com。